Chapter 2 ___________________________________________________________ Product Overview

VAISALA ________________________________________________________________________ 11

CHAPTER 2

PRODUCT OVERVIEW

This chapter introduces the features, advantages, and the product

nomenclature of the HMM100.

Introduction to HMM100 Series Humidity

Modules

The HMM100 series humidity measurement modules are open frame

modules intended for integration into environmental chambers and other

chambers where high temperature tolerance is important. The modules

provide a single analog output channel for relative humidity (RH) or

dewpoint (T

d

). Dewpoint is the recommended quantity if the

measurement environment has significant temperature differences.

Table 3

Quantities Measured by HMM100

Quantity

Range

Relative humidity (RH)

0 ... 100%

Dewpoint (Td)

-20 ... +100 ºC (-4 ... +212 ºF)

All HMM100 modules include automatic temperature compensation

across the operating temperature range using an integrated PT1000

sensor. No external temperature compensation should be applied to the

output.

HMM100 modules do not incorporate probe heating or chemical purge

functionality. If the measurement environment is prone to condensation,

or likely to cause chemical contamination of the sensor, other Vaisala

instruments may be more suitable to the application.

The cable and probe of the HMM100 are attached at the factory. They

should remain connected to the module at all times; do not cut and re-

connect the cable.