Set temporary pressure compensation value, Set temporary pressure – Vaisala HMM100 User Manual
Page 45

Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 43
NOTE
You can use the Vaisala Humidity Calculator to simulate the effect of
pressure change to dewpoint. The Humidity Calculator can be found at:
Set Temporary Pressure Compensation
Value
Use the XPRES command to show or set the temporary ambient pressure
value for pressure compensation. This value is cleared to 0.0 at reset.
XPRES [x.xxxx]
where
x.xxxx
= Absolute pressure in the measuring environment.
The pressure unit is bar.
NOTE
Remember to save the changes!
Example:
>xpres
Pressure (bar) : 0
>
>xpres 1.029
Pressure (bar) : 1.029
>save
Saving settings...done
>