Calibrate Analog Output

Use the ACAL command to calibrate the currently active analog output

mode. You must use this command if you change the analog output mode

using the AMODE command.

ACAL

After you give this command, the analog output of the module is set to

the low limit of the selected output (voltage or current). Measure the

output and enter the measured value. After entering the low limit, you

must measure and enter the high limit also.

Example (analog output in 4 ... 20 mA mode):

>acal

I1 (mA) ? 3.981

I2 (mA) ? 20.002

2.00985160E+03 a0

1.12789450E+03 a1

0.00000000E+00 a2

OK

>

Example (analog output in 0 ... 1 V mode):

>acal

V1 (V) ? 0

V2 (V) ? 0.998

5.45999950E+02 a0

1.64128260E+03 a1

0.00000000E+00 a2

OK

>