Configuring analog output, Set analog output mode – Vaisala HMM100 User Manual
Page 46
User's Guide _______________________________________________________________________
44 ___________________________________________________________________ M211024EN-C
Configuring Analog Output
Set Analog Output Mode
Use the AMODE command to show or set the operation mode of the
analog output. This command is only useful with the 3-wire output
module, as the output on the 2-wire model is always 4 ... 20 mA (loop
powered).
NOTE
If you change the analog output mode, you must calibrate the new mode
using the ACAL command. See section Calibrate Analog Output on page
AMODE [mode]
where
mode = Analog output mode. Modes for 3-wire output model:
0..1V
0..5V
0..10V
0..20mA
4..20mA
Modes for 2-wire output model:
4..20mA
NOTE
Remember to save the changes!
Example (show current mode):
>amode
Ch1 : 0..10V
>save
Saving settings...done
>
Example (set mode to 0...20 mA):
>amode 0..20ma
Ch1 : 0..20MA
>save
Saving settings...done
>