Device information and status, Show device information, Table 7 – Vaisala HMM100 User Manual
Page 33: Advanced serial commands

Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 31
Table 7
Advanced Serial Commands
Command
Description
ACAL
Calibrate analog output
AMODE
Set/show analog output mode
ASEL [RH/TD]
Set/show analog output quantities and
scaling
ATEST
Test analog output; forces the output ot the
given value
CALINFO
Set/show calibration info
CRH
Calibrate humidity measurement after
CT
Calibrate temperature measurement
FRESTORE
Restore all module settings to factory
defaults
L
Show user adjustment parameters
LI
Set user adjustment parameters
SMODE
[STOP/RUN/POLL] Set the serial interface mode
ZERO
Set correction by trimmers to zero
Device Information and Status
Show Device Information
The ? command outputs a listing of device information.
?
Example:
>?
Device Name : HMM100
SW Name : HM100
SW model : VAISALA
SW version : 1.1.10.3031
Serial number :
Calibrated :
Address : 0
Operating mode : 3W
Status : ON
>
If the module is in poll mode, but a connection has not been opened using
the OPEN command, issue the ?? command. For a description of the
serial interface modes, see section Set Serial Interface Mode on page 41.
??