Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 35

Calibration Commands

Commands that have to do with calibration on the serial line are

described below. For the calibration procedures, see section Calibration

on page 53.

Calibrate Temperature Measurement

Use the CT command to perform a one-point or two-point temperature

(T) calibration.

CT

When performing a one-point calibration, you need to place the probe in

a single temperature reference. Run the command and enter the exact

temperature of the reference after the measurement has stabilized. Exit

the command by pressing the ESC key before proceeding to the second

point (see example 1 below). To update the value, press ENTER without

inputting a value.

For two-point calibration, the second reference point must be at least

20 ºC warmer than the first. The measurement reading for the second

point will appear when this requirement is met.

NOTE

Remember to save the changes!

Example: one-point calibration

>ct

T 21.9827 Ref1 ? 22

Press any key when ready

OK

>

Example: two-point calibration

>ct

T 22.0007 Ref1 ? 22

Press any key when ready ...

T 44.9847 Ref2 ? 45

OK

>