Chapter 5 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 57

8.

Give the CT command. The output from the command shows the

currently measured T value, and prompts you to enter the value of

the reference 1. This is the reference on the colder end, which

produces the offset correction for the measurement.

9.

You now have three choices:

- To apply the correction, type the temperature of the reference in

degrees celcius and press ENTER.

- To exit the CT command without adjusting the measurement, press

the ESC key and exit this procedure.

- To refresh the measured value without typing in the correction,

press ENTER.

10. If you only want to perform a one-point calibration (offset

correction), press the ESC key after entering the correction. The

CT command will exit, and you can move to step 14.

>ct

T 21.9847 Ref1 ? 22

Press any key when ready

OK

>

To continue with the calibration of the warmer end (gain

correction), press ENTER.

11. Insert the probe in the second temperature refence and wait for 20 –

40 minutes for the reading to stabilize. The measurement reading

for the second point will appear when the measured temperature is

at least 20 ºC warmer than the first reference point.

12. Type the temperature of the second reference point and press

ENTER to apply the correction.

T 44.9847 Ref2 ? 45

OK

>

The CT command is now finished and both adjustments have been

applied.

13. Use the L command to verify that the new user adjustment

parameters are in use:

>l

14. Measure the value of the analog output to verify that the

measurement is OK.