Select analog output quantity and scaling, Test the analog output – Vaisala HMM100 User Manual
Page 47
Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 45
Select Analog Output Quantity and
Scaling
Use the ASEL command to show or select the quantity output by the
analog channels, and the scaling that is used. If you do not enter scaling
limits, default values are used.
ASEL [xx] [low high]
Where
xx = Analog channel output quantity
low = Low limit of scaling
high = High limit of scaling
NOTE
Remember to save the changes!
Examples:
>asel
Quantity : RH
Level Lo : 0
Level Hi : 100
>save
Saving settings...done
>
>asel td
Quantity : TD
Level Lo : -40
Level Hi : 60
>save
Saving settings...done
>
>asel td -10 60
Quantity : TD
Level Lo : -10
Level Hi : 60
>save
Saving settings...done
>
Test the Analog Output
Use the ATEST command to test the analog output. The ATEST
command will force the output to the given value, which can then be
measured with a calibrated multimeter. The command uses the currently