Set Output Interval

Use the INTV command to show or set the output interval of the serial

line measurement messages (applies when R command or RUN mode is

used). The shortest output interval is one second. This command has no

effect on the operation of the analog output.

NOTE

Remember to save the changes!

INTV [n xxx]

where
n

= Time interval in range 1 ... 255

xxx

= time unit = "S", "MIN", or "H"

Example:

>intv 1 s

Value : 1

Unit : S

>save

Saving settings...done

>

Output a Reading Once

Use the SEND command to output a single measurement reading from

the module.

SEND [aa]

where
aa = Module address (0 ... 99). Must be specified if the module is

in POLL mode, and a line has not been opened using the

OPEN command.

Example:

>send

RH= 25.12 % T= 24.91 'C

>