The polling mode can be used when more than one module is connected

to one serial bus. When set to POLL state, the modules communicate one

at a time when the specific module address is called on the serial line.

See descriptions of the commands ADDR on page 41 and OPEN on

page 48.

Examples:

>smode stop

Output mode : STOP

>

>smode run

Output mode : RUN

>

>smode poll

Output mode : POLL

>

Configuring Measurement Parameters

Set Environmental Compensation Values

Use the ENV command to show or set the environmental compensation

values. In HMM100, the humidity measurement can be compensated for

ambient pressure. The value set using ENV command is a permanent

value which will survive a restart. Use the command XPRES instead if

the value is changed frequently; see section Set Temporary Pressure

Compensation Value on page 43.

ENV [x.xxxx]

where
x.xxxx

= Absolute pressure in the measuring environment.

The default is 1.013. The pressure unit is bar.

Example:

>env

Pressure (bar) : 1.013

>

>env 0.980

Pressure (bar) : 0.98

>save

Saving settings...done

>