Badger Meter M-Series M2000 User Manual
Page 58

Menu Manager Configuration Errors
Error
Description
Recommended Action
140
Output 3: Configuration Error – Full
scale frequency exceeds limits of
relay (1000 Hz)
Reduce full scale frequency setting of output when hardware is configured for relay operation
150
Output 3: Configuration Error – Full
scale frequency exceeds limits (10
kHz)
Reduce full scale frequency setting of output when hardware is configured for open collector
operation
170
Output 1/2: Output Type
Configuration Error
This error is observed when the function type is 24V DC and the output type is changed from
Normally Open to Normally Closed It is required for 24V DC output operation that the output type be
Normally Open
171
Output 1/2: Output Type
Configuration Error
This error is observed when the function type is ADE and the output type is changed from Normally
Open to Normally Closed It is required for ADE operation that the output type be Normally Open
190
Full Scale Flow: Entered Value
exceeds limits
Value entered exceeds the absolute maximum flow the meter supports Reduce the value for this
parameter or consider increasing pipe diameter
Display Error / Status Messages
Error Message
Possible Cause
Recommended Action
Err: Detector
No detector connection with amplifier
Check detector and cable connections in accordance with this manual
Connection between amplifier and detector
Contact Technical Support
Supply voltage too low
Contact Technical Support
Grounded coils in meter
Contact Technical Support
Water in detector
Contact Technical Support
Err: Empty pipe
Pipe may not be full
Make sure all trapped air is out of system
If fluid or fluid conductivity recalibrate the parameter
Err: Full scale
Actual flow rate is exceeding programmed flow Reduce flow rate or increase the programmed full scale value by more than 5%
Err: AD-Range
AD-Converter is exceeding signal limits
Check the grounding scheme of the meter installation See grounding section in this
manual Verify pipe is not empty
Err: AD-INT
Initialization of AD-Converter unsuccessful
Contact Technical Support
Err: Rollover
Rollover counters have exceeded limit
Clear all totalizers
Err: Rollover Status
Totalizer rollover has occurred
Reload totalizer then clear all totalizers
Err: Simulation
I/O simulator is enabled
Disable simulator in I/O menu
M-Series® M2000 Electromagnetic Flow Meter
Page 58
August 2012