Partially-filled pipe situations, Wrong right, Wrong – Badger Meter M-Series M2000 User Manual
Page 12: Right
Partially-Filled Pipe Situations
In some locations, the process pipe may be momentarily only partially filled Examples include: lack of back pressure,
insufficient line pressure and gravity flow applications
To eliminate these situations, do not install the meter at the highest point of the pipeline
WRONG
FL
OW
Figure 12: Incorrect Meter Placement
Do not install the meter in a vertical, downward flow section of pipe
Always position the ON/OFF valves on the downstream side of the meter
WRONG
RIGHT
FL
OW
FL
OW
FL
OW
WRONG
FL
OW
RIGHT
WRONG
RIGHT
FL
OW
FL
OW
FL
OW
WRONG
FL
OW
RIGHT
Do not install in a vertical, downward position
Position "On/Off" valves on downstream side
Figure 13: Position Valves on Downstream Side
To minimize the possibility of partially-full pipe flows in horizontal, gravity or low pressure applications, create a pipe
arrangement that ensures the detector remains full of liquid at all times
Figure 14: Pipe Positioned to Keep Water in Detector
M-Series® M2000 Electromagnetic Flow Meter
Page 12
August 2012