Advanced Features
Empty Pipe Cal.
[Default]
S
Fluid conductivity impacts the performance of empty pipe measurements If you require empty pipe
detection, you should perform this empty pipe calibration procedure
Before starting the empty pipe calibration, verify that empty pipe detection is enabled Also, run both
the empty pipe and the full pipe calibration procedures
Calibrating an Empty Pipe
Before calibrating an empty pipe, verify that the pipe is empty
To calibrate with an empty pipe, follow these steps from the Advanced menu:
1 Select Empty Pipe Cal to view the Calibration menus
2 Select Cal. Empty Pipe to view the Empty Pipe Calibrate menu
3 To enable calibration, place the cursor on the calibration enable line and press [E]
EMPTY PIPE CALIBRATE
Volts = 3.00
> Cal [ON] E=OFF
Exit with Save
FULL PIPE CALIBRATE
Volts = 1.515
> Cal [OFF] E=ON
Exit with Save
4 Wait 30 seconds for voltage measurement to stabilize
5 To save the setting, place the cursor on Exit with Save and press [E]
Calibrating a Full Pipe
Before calibrating a full pipe, verify that the pipe is full
To calibrate with a full pipe, follow these steps from the Advanced menu:
1 Select Empty Pipe Cal to view the Calibration menus
2 Select Cal. Full Pipe to view the Full Pipe Calibrate menu
3 Enable calibration by placing the cursor on the calibration enable line and press [E]
EMPTY PIPE CALIBRATE
Volts = 3.00
> Cal [ON] E=OFF
Exit with Save
FULL PIPE CALIBRATE
Volts = 1.515
> Cal [OFF] E=ON
Exit with Save
4 Wait 30 seconds for voltage measurement to stabilize
5 To save the setting, place the cursor on Exit with Save and press [E]