Badger Meter M-Series M2000 User Manual
Page 39
Clear Totals
T1
S
The uni-directional totalizer is reset within the menu manager or through remote communications
Clearing T1 also clears the associated rollover counter
T2
U
The uni-directional totalizer is reset within the menu manager, through remote communications or with
properly-configured digital input (function = remote reset) Clearing T2 also clears the associated rollover
counter
T+
S
The bi-directional forward flow totalizer is reset within the menu manager or through remote
communications Clearing T+ also clears the associated rollover counter
T–
S
The bi-directional reverse flow totalizer is reset within the menu manager or through remote
communications Clearing T– also clears the associated rollover counter
Tn
S
The bi-directional net totalizer, when reset, clears both the forward and the reverse flow totalizers (T+ and
T–) This is reset within the menu manager or through remote communications Clearing Tn also clears the
associated rollover counter
PS
S
The batch totalizer is reset to the configured preset amount value It is reset within the menu manager,
remote communications or through a properly-configured digital input (function = batch reset)
Tpwroff
S
The totalizer accumulating meter time without external power is reset with the menu manager or through
remote communications
Communication: Port Settings
Interface
[MODBUS RTU]
S
The Interface parameter lets you configure how the RS232 communication port will be used
• MODBUS RTU
• Remote menu (RDI – Remote Display Interface)
• Primo 3 x
• Disable port
The remote menu interface will check for display updates once a second If a change is detected, the
display contents will be transmitted in ASCII format over the RS232 communication port The remote
menu interface also allows for menu navigation and control of the meter as if using the external push
buttons Keyboard control characters such as
menus
The Primo 3 x interface will emulate the legacy Primo 3 x Protocol This protocol will transmit an ASCII
string in the following format every 500 ms:
“RATE;0 0000; GPM; TOT1;150 0000; USG ; TOT2;150 0000; USG ;” – For Unidirectional Mode
“RATE;0 0000; GPM; TOT+;10 0000; USG ; TOT-;50 0000; USG ;” – For Bidirectional Mode
To change the Interface follow these steps from the Port A Settings menu:
1 Select Interface to view the Interface display
2 Select the desired interface
3 Press [E] to save and return to the Port A Settings menu
Installation & Operation Manual
Page 39
August 2012