Protocol Type
The protocol type enables the ADE interface
Selecting S1 or S2 will automatically configure
the Digital Input and Digital Output #1 for ADE
operation Manually configuring the input and
output for ADE operation is not allowed and will
result in a on-screen error
See "ADE Interface" on page 52 for further
details
S1 – Standard ADE protocol
S2 – Enhanced ADE protocol, provides additional
digital information
Disabled – disables and removed ADE
configuration
Dial Type
4 – 8 dials
Analog Calibrate
Custom Settings
[Zero Scale: 0 mA]
[Full Scale: 0 mA]
S
To set the analog calibration custom settings, follow these steps from the
Advanced menu:
1 Select Analog Calibrate to view the Analog Calibrate menu
2 Select Custom Settings to view the Custom Settings display
3 Select one of the following:
• Offset 4 mA
• Offset 20 mA
4 Configure desired offset
5 Press [E] to save the option and to return to the Custom Settings menu
6 Press [E] to return to the Analog Calibrate menu
Factory Settings
[Factory Set]
S
To change the analog calibration factory settings, follow these steps from the
Advanced menu:
1 Select Analog Calibrate to view the Analog Calibrate menu
2 Select Factory Settings to view the Factory Settings display
3 Select one of the following:
• Calibration Point A
• Calibration Point B
4 Set the calibration point to the measured output current
5 Press [E] to save the option and to return to the Factory Settings menu
6 Press [E] to return to the Analog Calibrate menu