Badger Meter M-Series M2000 User Manual
Installation & operation manual, M-series® m2000
M-Series® M2000
Store/Restore Feature
Installation & Operation Manual
IOM-194-01-EN (August 2012)
53400-194 Rev. 1
IMPORTANT
The Store/Restore feature requires firmware version 1.11 or
later. Reference Badger Meter P/N 67354-003 to obtain a
firmware upgrade kit.
DESCRIPTION
The Store/Restore feature is intended to save installation
costs and reduce installation time. This feature is also
intended to protect meter configuration and assure the
operator that the meter is properly configured. Over time
and handling of the meter, the meter configuration could
change. The Store/Restore feature allows the meter to be
quickly set to the operator’s original configuration.
PERFORMANCE
The Store/Restore feature provides the ability to save meter
configuration to a memory token. The memory token can
be placed into alternate meters so that each meter in the
system has the same configuration. The memory token can
also be stored in a safe place and later used to restore meter
configuration if the configuration of the meter is in question.
The operator has the option to select which groups of
configuration parameters are stored to the memory token,
providing additional flexibility for system installation.
OPERATION
The means to enable and configure the Store/Restore
feature are found under the Advanced menu of the menu
manager located in the Token Copy submenu. The menu
items of Token Copy are:
• Configuration
Token Properties
Parameter Selection
• Store to Token
• Restore to Token
The Configuration submenu allows the operator to configure
token properties and select the groups of parameters that
are to be stored to the token. Token properties and the
parameter selection configurations should be configured
first, prior to copying meter information to the
memory token.
Token properties allow the token to be configured in a
way that the behavior of the token can be defined. Such
properties include:
• Write Protect – Protects the token from being overwritten
• Power on Load – Requests the meter to load the contents
of the token into the meter at power on.
• One Time Load – Requests the meter to erase the token
after the contents of the token have been stored to the
meter.
Write protect cannot be undone by the operator. Caution
should be used when configuring the token for write-
protect mode.
The chart below lists the categories of parameters that can
be individually selected for storing to the memory token.
Parameter Categories
Calibration
Pipe diameter
Detector factor
Detector offset
Communication port A/B
Interface (for port A only)
Node address
Baud rate
Parity
Date bits
Stop bits
Extended node address (for
port B, Profibus DP only)
Security
Admin pin
Service pin
User pin
User
All other meter parameters
including measurement and
I/O settings
Nontransferable parameters include:
• Serial Number
• Excitation Frequency
• Scale Factor
• Analog output calibration factors
(including custom offsets)