2.4.0 FREQ. OUTPUT

2.4.1 TEST

To check the functioning of both the frequency output and
the externally connected instruments and wiring, the frequen-
cy output can be set to 5 specific values:

1, 10, 100, 1000 and 10000 Hz

The value shown on the display should match the indication
on the instrument connected to the frequency output.

Press to start the test and use the

key to select output frequency.

Press to exit the function.

73