Altosonic – KROHNE UFM 600T EN User Manual

Page 67

background image

Fig.D.9: Positive flow (F)

3.4.2 PULSOUTP.

:

Frequency output mode

Select one of the two frequency output modes using the

key and press to store:

PULSE/UNIT

:

Pulses per unit of volu-
me.

PULSRATE

:

Pulses per unit of time
at full scale flow.

Example of PULSE/UNIT

:

Pulse value

:

10 pulses per m3 (set
via Fct.3.4.3)

After 60 m3 have passed

:

Output: 600 pulses
since start.

Example PULSRATE

:

Full scale setting

:

1000 liters per second
(set via Start Up Menu)

Pulse rate full scale flow

:

1000 pulses per second
(set via Fct. 3.4.3)

At 600 liters per second

:

Output: 600 Hz signal.

64

ALTOSONIC

Programming mode

UFM 600 T