beautypg.com

KROHNE UFM 600T EN User Manual

Page 40

background image

Preparation for new measurements
Providing any previous flow data can be discarded, a data
collection session can be started as follows:
1

Select the STORE or BOTH option of function 3.6.3 in the
programming mode

2

Clear the RAM by means of one of the two ways to Reset
Data

3

Start the measuring mode

4

If applicable reset the totalizers

Preparations for continued measurements
It is advisable not to activate the STORE option of function
3.6.3 in the programming mode until the actual measuring
period is to be started; otherwise unnecessary data blocks
are being stored and storage space is lost.

After installing the transducers it is advisable to run in the
measuring mode for some time during which the data is only
sent to the display (function 3.6.3 OFF).
The readings can be checked and adjustments made (e.g.
adjust full scale value) to ensure that the suitable data is col-
lected.

The actual measuring period with data storage (and the RAM
partially filled) can then be started as follows:

1

Select the STORE or BOTH option of function 3.6.3 in the
programming mode

2

Leave the programming mode and enter the measuring
mode via the DIAMETER prompt of the Start Up menu

3

If applicable, reset the totalizers

39