Trouble Shooting

1

The flowmeter does not respond after connection to
the main power supply
(no display or output messages).

- Check fuse.

2

The error marker * appears on the display

-

Leave the measuring mode and enter the Quit/Reset
menu
(press ; see section C.5).
Cure the cause if necessary, and select
ERR.ACQUIT to reset the error marker.

3

Low signal marker: 3 or 4 markers on.

Cause

Solution

Incorrect transducer

Reposition transducers (see sec-
tion position C.1.3) using sufficient
coupling grease and check the
transducer distance afterwards.

Poor sonic coupling

Check the pipe wall surface;

between pipewall and

remove any rust, dirt and/or thick

transducer surface

layers of paint or coating.
Apply sufficient grease on the
transducers and make sure not to
wipe off any grease from the trans-
ducers while they are being rein-
stalled.

Pipe is not completely

Fill the pipe completely or select a

filled

pipe section that is always com-
pletely filled with liquid.

Broken signal cable

Use a multimeter to check the
condition of the cables and the
connectors. See your supplier for

46

ALTOSONIC

Operation

UFM 600 T