NOTE:

-

The cable connections at the transducers must always
point away from each other (fig. C.5).

-

Make sure that the transducers are not placed over a
welding seam.

Fig. C.5 : Upstream and downstream connection

5

Use the cm scale on the mounting rails to position
the transducers at the proper distance from each other
and hand tighten the securing screws.

6

Check using the indication of the low signal marker (see
C 3.1) whether the unit is functioning properly.

Continue with Zero Set function (C.1.4.)

NOTE: REMOVAL OF THE TRANSDUCERS FROM

PIPE WALL

When removing the transducers from the pipe wall, especially
after they have been in place for a long period, care should
be taken:
• Not to damage the tranducers by pulling strongly at the

securing screw.

• Therefore, to remove a tranducer, it must be removed in

a slicing motion. Do not just pull it straight off the pipe all !!

sensor distance S

FLOWDIRECTION