Altosonic – KROHNE UFM 600T EN User Manual

Page 37

background image

C.3.2

Error messages

If the microprocessor produces an error message a * sign
appears at the lower left hand side of the display.
Enter the QUIT / RESET mode to view the error message(s)
(see chapter C.5).
It depends on the settings of the Installation functions 3.2.4
and 3.2.6, whether the error messages will be displayed in
between the flow data.

List of Error Messages:

SIGN LOST

:

Transducer disconnected

TOTALIZER

:

Overflow of the totalizer.

FREQ.OUTP. F:

Overflow of the frequency signal; full
scale value is too small

CUR. OUTP. I :

Overflow of the current signal; full
scale value is too small

CAL. DATA

:

The transducer calibration is
distorted; contact your supplier

EE1 EE2

:

Error detected during memory check;
switch unit off and on; if message still
reappears after two attempts, contact
your supplier

EEPROM2

:

See EE1 EE2

ROM

:

Checksum error detected during
memory test; contact your supplier

RAM

:

Error detected during verification of
parameter settings stored previously;
use function 3.5.8 DEFAULT SET and
switch off and on; if message
reappears contact your supplier.

36

ALTOSONIC

Operation

UFM 600 T