2.0.0 TEST

2.1.0 DISPLAY

2.1.1 DISPLAY TEST

All individual segments of the liquid crystal display are
activated in a sequence ending in a flashing display of all
segments.
No action can be taken.
Select YES and press to start the test.

2.2.0 PROCESSOR

The microprocessor performs a self test.
No action can be taken.

Select YES and press to start the test.

If the test does not end with the message NO ERROR
contact your dealer.
Press to exit the function.

2.3.0 CURR. OUTP. I

2.3.1 TEST

To check the functioning of both the analog output and the
externally connected instruments and wiring, the analog out-
put can be set to 7 specific values:

0, 4, 8, 12, 16, 20 and 22 mA

The value shown on the display should match the indication
on the instrument connected to the current output.

Press to start the test and use the

key to select output current.

Press to exit the function.