beautypg.com

Technical data, 2 technical data – KROHNE OPTISWIRL 4070 EN User Manual

Page 70

background image

8

TECHNICAL DATA

70

OPTISWIRL 4070

www.krohne.com

02/2013 - 4000150606 MA OPTISWIRL4070C R07 en

8.2 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Application range

Flow measurement of liquids, gases and vapours

Measuring principle

Karman vortex street

Measured value

Primary measured value

Number of separated vortices

Secondary measured value

Operating and standard volumetric flow, mass flow

Signal converter

Versions

Compact

Remote version - cable length ≤ 15m (49 ft)

Option: Ex version

Measuring sensor

Standard

Flange version (with integrated temperature measurement), sensor: F

Sandwich version (with integrated temperature measurement), sensor: S

Optional

Basic device with additional pressure measurement

Basic device with additional pressure measurement and shut-off valve for pressure

sensor

Dual measuring device in both flange and sandwich version (redundant

measurement)

Dual measuring device in flange version with additional pressure measurement

Flange version with simple reduction, sensor: FR

Flange version with double reduction, sensor: F2R

Display and user interface

Local display

2 lines, 10 characters per line

Operating and display languages

German, English, French

MA_OPTISWIRL4070C_R07_en_150606_PRT.book Page 70 Wednesday, February 13, 2013 2:47 PM