6.5.5 Menu item Tests

Every current value has to be confirmed with ^. Only after this is the current output
set to the respective value. The following is then displayed:

• "Contin. Yes":

"Contin. Yes":

"Contin. Yes":

"Contin. Yes": continue with the test of current output function

• "Contin. No":

"Contin. No":

"Contin. No":

"Contin. No": back to menu

The test current is maintained until the menu has been exited.

Level

Designation

Selection / entry

Explanation

2.1.1

Test I

4 mA ^

Test current output

8 mA ^

12 mA ^

16 mA ^

20 mA ^

2.1.2

Test P

0.5003 Hz ^

Test pulse output

Table 6-10: Menu item Test