ELECTRICAL CONNECTIONS

40

OPTISWIRL 4070

www.krohne.com

02/2013 - 4000150606 MA OPTISWIRL4070C R07 en

4.6 Protection category

The measuring device meets all requirements of protection category IP66/67.

Therefore it is essential to observe the following points:

• Use only original gaskets. They must be clean and free of any damage. Defective gaskets must

be replaced.

• The electrical cables used must be undamaged and must comply with regulations.
• The cables must be laid with a loop 1 upstream of the measuring device to prevent water

from getting into the housing.

• The cable feedthroughs 2 must be tightened.
• Align the measuring device so that the cable feedthrough is never facing up 3.
• Close the unused cable feedthroughs using blind plugs 4.
• Do not remove the required cable bushing from the cable feedthrough.

CAUTION!
After all servicing and maintenance work on the device, the specified protection category has to
be ensured again.

Figure 4-5: Cable feedthrough