TECHNICAL DATA

8

69

OPTISWIRL 4070

www.krohne.com

02/2013 - 4000150606 MA OPTISWIRL4070C R07 en

Technical data

8.1 Functional principle

Vortex flowmeters are used to measure the volumetric flow of gases, vapours and liquids at
completly filled pipes.
The measuring principle is based on the Karman vortex street. The measuring tube contains a
bluff body, behind which vortex shedding occurs. The frequency ffff of the vortex shedding is
proportional to the flow rate vvvv. The nondimensional Strouhal number S

S

S

S describes the

relationship between vortex frequency ffff, width b

bb

b of bluff body and the mean flow velocity vvvv:

The vortex frequency is recorded at the sensor and evaluated at the converter.

Figure 8-1: Functional principle

=

f

S . v

b