ELECTRICAL CONNECTIONS

38

OPTISWIRL 4070

www.krohne.com

02/2013 - 4000150606 MA OPTISWIRL4070C R07 en

Connection of shielding connection cable remote version

4.5 Grounding connections

The grounding can be done either by connecting the PE terminal in the housing or the PE
terminal on the connection piece between measuring sensor and signal converter. Both of these
electrical connections are equally effective from a technical point of view.

CAUTION!
The cable shielding is connected only on one side under the U-clamp terminal 1 in the wall
mount bracket connection box.

Grounding connection compact version

1 Electrical grounding connection on connection piece between measuring sensor and signal converter.
2 Electrical grounding connection on housing

CAUTION!
The measuring device has to be grounded properly to achieve accurate measurement. The
grounding wire may not transfer any interference voltage.
Do not use this grounding wire to ground any other electrical devices.