Operation – KROHNE OPTISWIRL 4070 EN User Manual
Page 55
OPERATION
6
55
OPTISWIRL 4070
www.krohne.com
02/2013 - 4000150606 MA OPTISWIRL4070C R07 en
(available if Maes.Inst =
Mass, see menu item
1.1.3)
kg
Unit ↑...^
Select totalizer unit for mass flow measurement
0000000000
kg ↑…↑...↑...^
Enter preset totalizer value
Reset Yes /
Reset No ↑...^
Reset totalizer /
Do not reset totalizer/
Disp. off /
Disp. on ↑...^
Display totalizer /
Do not display totalizer
3.3
HART settings
3.3.1
Poll. Adr.
→
000
0 – 15
↑…→↑...→↑...^
Enter HART polling address for "Multidrop operation"
3.3.2
HART SV
→
Total Flow ^
HART secondary variable
3.3.3
HART TV
→
Temp. ^
HART tertiary variable
Density ^
3.3.4
HART 4V
→
Density ^
HART quaternary variable
Temp. ^
3.4
Set fluid and medium
3.4.1
Fluid 1
→
Set fluid type
Gas ↑...
Gas
Liquid. ↑...
Liquid
Steam ↑...^
Steam
3.4.2
Medium
→
Set medium
(available if Fluid = Gas or
Liquid, see menu item
3.4.1)
Custom ↑...
Customer medium
(available if Fluid = Steam,
see menu item 3.4.1)
Sat.Steam ↑...^
Saturated steam
Custom
1 Density check
3.5
Set pressure, temperature and density
3.5.1
T-sensor
→
Internal temperature sensor
No ↑...
No temperature sensor available
Yes ↑...^
Temperature sensor available
3.5.3
Sat.P/T
→
(available if Fluid = Steam,
see menu item 3.4.1)
Density calculation for saturated steam
Sat. T ^
Density calculation for saturated steam with the
saturated temperature
3.5.4
Oper. Temp.
→
Operating temperature
°
C
Unit ↑...^
Set temperature unit
0000000.0
°
C ^
Operating temperature
Disp. on /
Disp. off ↑...^
Display temperature /
Do not display temperature
Version Basic
Level
Designation
Selection / entry
Explanation
MA_OPTISWIRL4070C_R07_en_150606_PRT.book Page 55 Wednesday, February 13, 2013 2:47 PM