INSTALLATION

3

27

OPTISWIRL 4070

www.krohne.com

02/2013 - 4000150606 MA OPTISWIRL4070C R07 en

3.5 Inlet and outlet runs

3.5.1 Minimum inlet runs

1 General inlet run without disturbing flow ≥ 20 DN
2 Behind a control valve ≥ 50 DN
3 After a pipe diameter reduction ≥ 20 DN

1 After a single bend 90° ≥ 20 DN
2 After a double bend 2x90° ≥ 30 DN
3 After a double three-dimensional bend 2x90° ≥ 40 DN