GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

64

Communication Mode

The same baud rate and data format must be set to the instrument and the computer.

The baud rate of the RS-232 interface can be set as listed in the following table.

300

Baud

900

Baud

1200

Baud

2400

Baud

4800

Baud

9600

Baud

The data transmission format is N-8-1 (no parity bit, 8 data bits, 1 stop bits).

Computer’s Connection

A personal computer with a COM port is the essential facility in order to operate the

instruction via RS232 interface.

The connections between GRS-6052A/6032A and computer are as follows:

1) Connect one end of a RS232 cable to the computer.

2) Connect the other end of the cable to the RS232 port on the GRS-

6052A/6032A.

3) Turn on the GRS-6052A/6032A.

4) Turn on the computer.

GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

65

The RS232 connection testing

If you want to test whether the RS232 connection is working or not, you can send a

command from computer. For instance, using a terminal program send the query

command (uppercase)

*IDN?

should return the Manufacturer, model number, serial number and firmware version

in the following format:

GW,GRS60X2A,V.1.10

If you do not receive a proper response from the GRS-60X2A, please check if the

power is on, the RS232 configurations are the same on both sides, and all cable

connections are active.