GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

52

Figure 5-16

ROLL mode

The displayed waveform is rolled from right to left (0.2s/DIV to 100s/DIV).

The right end of each trace is the updating point of a new data. The Roll mode

facilitates the measurement of a signal of approximately 100Hz or lower. Press

the STOP switch to stop the ROLL mode and hold the final waveform on the

CRT.

GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

53

Figure 5-17

NOTE:

Aliasing:

While measuring the signal in such STORAGE mode as SMPL, AVG, etc., the

aliasing can be occurred by inputting a signal of more than half of the

frequency with respect to the sample clock frequency at the sweeping range is

added.

When the aliasing is occurred, the waveform of the input signal frequency

minus the sample clock frequency will be displayed. It is possible that this

display is judged a current waveform. If the aliasing is suspected, select the

REAL TIME mode and check if the display is the same as that in the actual

operation mode.