beautypg.com

GW Instek GRS-6000A Series User Manual

Page 10

background image

GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

14

(3) INTEN—Control knob (REAL TIME Mode only)

The control knob is used for adjusting the traces intensity in the real time

mode. Turning the knob clockwise to increase the intensity while turning it

counterclockwise to decrease the intensity.

(4) FOCUS

The control knob effects both the trace and the readout sharply.

(5) CAL

The terminal provides a reference signal of 0.5Vp-p at 1kHz for probe

adjustment.

(6) Ground Socket—Banana Socket galvanically connected to safety earth

This socket can be used a reference potential connection for DC and low

frequency signal measurement purpose.

(7) TEXT/ILLUM—Control knob with a double function.

The pushbutton is for selecting the text readout intensity function or scale

illumination function, and indicates the letter “TEXT” or “ILLUM” in the

readout. Press the pushbutton for the following sequences:

“TEXT” — “ILLUM” — “TEXT”

The TEXT/ILLUM function is associated the VARIABLE (9) control knob.

Turning the knob clockwise to increase the text intensity or scale

illumination, while turning the knob counterclockwise to decrease it.

Pressing the knob to switch the TEXT/ILLUM on or off.

In the STORAGE mode, the brightness of the waveform on the screen can be

controlled by the “TEXT”.

(8) CURSORS MEASUREMENT FUNCTION

There are two pushbutton and associated the VARIABLE (9) control knob.

When the pushbutton is pressed, the three measurement functions will be

selected in the sequence as follows:

V—T—1/T—OFF

GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

15

V: Two horizontal cursors appear. The voltage between the two cursors is

calculated according to the setting of VOLTS/DIV, and displayed with

△V on the upper side of the CRT.

Single channel mode (CH1 or CH2):

The △V measuring result is automatically related to the deflection

coefficient of the active channel. The readout displays “△V1…” or

“△V2…”.

Dual channel mode:

The cursor lines must be set on the CH1 or CH2 signal. As the

deflection coefficients may be different, it will be required to select

between the deflection coefficient of CH1 and CH2.

ADD mode:

In ADD (addition) mode, normally two input signals are displayed as

one signal (sum or difference). As the result can only be determined if

both (calibrated) deflection coefficients are equal, the readout indicates

“ △ V…” without any additional channel information. Different

deflection coefficient settings or uncalibrated deflection coefficients

are indicated in the readout as “△V=…DIV”.

X-Y mode:

In the X-Y mode, the instrument is automatically set to △ V

measurement. The deflection coefficient selected for each channel may

be different, thus as in DUAL mode the △V cursor measurement

requires a channel selection. Under channel 1 (X signal) measuring

condition the cursor lines are displayed as vertical lines and the readout

displays “△VX…”. Pressing the pushbutton, select channel 2 (Y

signal) measuring, then the cursor lines are displayed as horizontal

lines and the readout indicates “△VY…”.

T: Two vertical cursors appear. The time between the two cursors is

calculated according to the setting of TIME/DIV, and displayed with