GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
30
(41) SINGLE
Pressing the pushbutton to set the SINGLE mode and the “SINGLE” message
will be indicated in the readout.
In this operation mode, a single signal acquisition process or sweep can be
started with a trigger. Providing the trigger circuit has been previously
activated with reset function, SINGLE is automatically switched to normal
triggering (NML LED lights up). Otherwise the trigger automatic would start
the signal acquisition processes without an input (trigger) signal.
Pressing the RUN/STOP pushbutton (reset function) again to resume a new
single event capture which then overwrites the previously recorded display.
(42) UTILITY
The instrument software contains several utility setting. Each time when the
UTILITY pushbutton is pressed, the readout displays the following message in
the sequence at the top right of the CRT:
RS232 baud rate
BEEP ON/OFF
FACTORY DEFAULT loading
RS232 Baud Rate
The setting of baud rate and data format on the instrument must be the same
as the one on the computer. The baud rate of the RS-232 interface can be
selected by turning the VARIABLE control knob according to the list as
follows:
300—900—1200—2400—4800—9600
Press the VARIABLE control knob to set RS-232 baud rate, the screen will
display “RM” in the upper left corner to show GRS-6052AA/6032 in the
Remote Control mode.
Note: When the baud rate is set, the front panel control will be locked.
Press UTILITY can unlock the front panel control and disable remote
control.
GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
31
BEEP ON/OFF
When the “BEEP” is displayed, turning the VARIABLE control knob to set
the beep on or off. In the “OFF” condition, the acoustic signals actuated by
the control limits are switched off.
FACTORY DEFAULT loading
When the “FACTORY DEFAULT” is displayed, pressing the VARIABLE
control knob to overwrite all panel setting memories (MEM0~MEM9), please
refer to the setting as follows:
REAL TIME mode
: ON
VERTICAL
: CH1: ON, CH2: ON
VOLTS/DIV: 0.5V
COUPLING: AC
HORIZONTAL
: TIME/DIV: 100μs
TRIGGER
: MODE: ATO
SOURCE: CH1
COUPLING: AC
SLOPE: