GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

30

(41) SINGLE

Pressing the pushbutton to set the SINGLE mode and the “SINGLE” message

will be indicated in the readout.

In this operation mode, a single signal acquisition process or sweep can be

started with a trigger. Providing the trigger circuit has been previously

activated with reset function, SINGLE is automatically switched to normal

triggering (NML LED lights up). Otherwise the trigger automatic would start

the signal acquisition processes without an input (trigger) signal.

Pressing the RUN/STOP pushbutton (reset function) again to resume a new

single event capture which then overwrites the previously recorded display.

(42) UTILITY

The instrument software contains several utility setting. Each time when the

UTILITY pushbutton is pressed, the readout displays the following message in

the sequence at the top right of the CRT:

RS232 baud rate

BEEP ON/OFF

FACTORY DEFAULT loading

RS232 Baud Rate

The setting of baud rate and data format on the instrument must be the same

as the one on the computer. The baud rate of the RS-232 interface can be

selected by turning the VARIABLE control knob according to the list as

follows:

300—900—1200—2400—4800—9600

Press the VARIABLE control knob to set RS-232 baud rate, the screen will

display “RM” in the upper left corner to show GRS-6052AA/6032 in the

Remote Control mode.

Note: When the baud rate is set, the front panel control will be locked.

Press UTILITY can unlock the front panel control and disable remote

control.

GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

31

BEEP ON/OFF

When the “BEEP” is displayed, turning the VARIABLE control knob to set

the beep on or off. In the “OFF” condition, the acoustic signals actuated by

the control limits are switched off.

FACTORY DEFAULT loading

When the “FACTORY DEFAULT” is displayed, pressing the VARIABLE

control knob to overwrite all panel setting memories (MEM0~MEM9), please

refer to the setting as follows:

REAL TIME mode

: ON

VERTICAL

: CH1: ON, CH2: ON

VOLTS/DIV: 0.5V

COUPLING: AC

HORIZONTAL

: TIME/DIV: 100μs

TRIGGER

: MODE: ATO

SOURCE: CH1

COUPLING: AC

SLOPE: