GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
18
Vertical controls
The vertical controls select the displayed signals and control the amplitude
characteristics.
(11) CH1—Pushbutton
(12) CH2—Pushbutton
Pressing briefly the CH1 (CH2) button to set the channel 1 (channel 2) of
the instrument on, the deflection coefficient will be displayed in the readout
indicating the current conditions.
(13) CH1 POSITION—Control knob
The vertical trace position of channel 1 can be set with the control knob. When
X-Y operation in the Storage mode, CH1 POSITION control knob is used for
the X deflection.
(14) CH2 POSITION—Control knob
The vertical trace position of channel 2 can be set with the control knob.
In X-Y operation, CH2 POSITION control knob is used for the Y
deflection.
GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
19
(15) ALT/CHOP
In the REAL TIME mode, the pushbutton has two functions, which are
required and available only when both channels are active.
ALT—Displays in the readout, indicates alternate channel switching. After
each time base sweeps the instrument internally, switches over from channel
1 and channel 2 and vice versa.
CHOP—Indicates chopper
The channel switching occurs constantly between channel 1 and channel 2
during each sweep.
In the STORAGE mode, ALT or CHOP mode is automatically selected by
TIME/DIV range. The ALT mode is established for the sweep range of
0.5ms/DIV or faster. The CHOP mode is established for the sweep range of
1ms/DIV or slower.
(16) ADD-INV—Pushbutton with double functions.
ADD– Displays the “+” symbol in the readout, indicates additional mode.
Whether the algebraic sum (addition) or the difference (subtraction) of both
input signals is displayed, depends on the phase relationship and the INV
setting. As a result, both signals are displayed as one signal. For correct
measurements, the deflection coefficients for both channels must be equal.
INV—Pressing and holding the pushbutton to set the channel 2 invert
function on or off. The invert on condition is indicated by the “ ” symbol
in the readout. The invert function causes the signal display of channel 2 to
be inverted by 180
o
.
(17) CH1 VOLTS/DIV
(18) CH2 VOLTS/DIV– Control knob for channel 1/channel 2 has double functions.
Turning the knob clockwise to increase the sensitivity in 1-2-5 sequence and
turning it in the opposite direction (CCW) to decrease. The available range is
from 1mV/DIV up to 20V/DIV. The knob is automatically switched inactive if
the related channel is switched off.