5-3.Adjustments and checks

Trace Rotation Adjustment

Normally, when the trace is in parallel with the center horizontal graticule line,

there will be no need to adjust the TRACE ROTATION. If necessary, adjust the

TRACE ROTATION to make the baseline trace parallel to the center horizontal

graticule line by using a small straight-blade screwdriver or alignment tool.

Probe Compensation

To minimize the distortion of measured waveforms, check the compensation of

your probes before using them. The probe compensation should be checked

periodically whenever the probes are moved to different input channels.

1. Install the probes onto the oscilloscope (Press the BNC connector onto the

channel input and rotate the connector to lock it into place).

2. Set the probe slide switches to the ×10 position.

3. Briefly pressing the CH1/CH2 button to set the oscilloscope to channel 1 and

channel 2.

4. Pressing and holding the P×10 button to set the indicated deflection

coefficient of the channel displayed in the readout as a symbol “P10”.

5. Attach the probe tips to the CAL connection in the front of the oscilloscope.

6. Set the oscilloscope controls to display both channels:

VERTICAL:

VOLTS/DIV

0.2V

COUPLING

DC

ALT/CHOP

CHOP

HORIZONTAL:

TIME/DIV

0.5ms

TRIGGER:

MODE

ATO

SOURCE

VERT

COUPLING

AC

SLOPE

7. Observe the displayed waveform and compare them with the waveforms

shown in figure 5-2. If either probe needs to be adjusted, proceed the step 8. If

GRS-6052A/6032A OSCILLOSCOPE

USER MANUAL

⎯ ⎯

39

either probe does not need to be adjusted, proceed the “Function Check”.

Figure 5-2 Typical Compensation Waveform

8. Adjust the probe by using a small insulated screwdriver. Slowly rotate the

adjustment control until the probe is properly compensated.