Viewing and Creating Setup Data
Memory Timeup
Specify the date and time to save the data in the internal memory to the external storage
medium.
• Timeup type
Time interval to save the data. Set [OFF] to disable the function.
Hour: 1 hour, Day: 1 day, Week: 1 week, Month: 1 month
• Date
Date (1 to 28) or the day of the week (SUN, MON, TUE, WED, THU, FRI, and SAT)
• Time (Hour)
Hours (00 to 23)
Channel (Setting the Burnout and RJC)
Click here (also selectable from [Setting] - [System Mode Setting]).
Set to the positive side (100%).
Copy and paste (see page 4-15).
Tools (see page 4-6)
Resets the range selection.
Channel number
Select the range by dragging & dropping the mouse.
Set the reference junction
compensation to internal or external.
Set to the negative side (0%).
Burnout
For thermocouple (TC) inputs, select how the measurement results are to be handled
when the thermocouple burns out.
RJC Volt (uV)
Internal: Use the DXP’s reference junction compensation function
External: Use an external reference junction compensation function.
Volt(uV): When the type is set to [External], set the compensation value in the range from
–20000 to 20000.
Copying and Pasting Setup Data
The items checked in [Copy Details] can be copied and pasted. For the operating
procedure, see page 2-5.
2.6 Configuring the System Mode