2-18
IM04L41B01-65EN
Batch
Click here (also selectable from [Setting] - [Engineering Mode Setting]).
Automatically increase the lot number by 1 at Memory Stop.
Use the lot number indication.
Header 1 to 3
Enter character strings (up to 64 characters) that are written to the batch and continuous
data files.
Calibration Correction
Can only be specified on measurement channels.
Click here (also selectable from [Setting] - [Engineering Mode Setting]).
Set point number
Enter values.
Select the number of set
points.
Select channel.
Number of Set Points
Select the number of set points from [OFF] or [2] to [16].
Select [OFF] if you are not using the correction function.
Selectable Range of Measured and True Values
Channels that are set to scaling or square root computation: –30000 to 30000 (the
decimal point position depends on the channel setting)
All other channels: Selectable range in the specified range
For the measured value, set a value that is greater than the previous measured value. If
you enter a setting that does not meet these conditions, the [Number of Set Points] is set
to [OFF] when you store the setup data, print the setup data, or send the setup data to
the DXP using the Communicator.
2.5 Confiuring the Engineering Mode