Teledyne LeCroy WaveJet 300A Series Remote Control User Manual
Page 56
WaveJet 300A Series
56
917976 Rev A
Vertical VDIV
Command/Query
DESCRIPTION
The VDIV command sets the vertical sensitivity of the
specified input channel in units of "V/div". When the
specified value exceeds the allowable range, it is
rounded according to the 1-2-5 step rule. The VDIV?
query returns the defined vertical sensitivity. While
the numerical value expressed in voltage is returned,
the suffix is not returned.
COMMAND SYNTAX
: = {C1, C2, M1} (WJ352/332/322/312)
The
used, and the suffix is also valid. For rounding rules,
see "Remarks".
QUERY SYNTAX
RESPONSE FORMAT
The currently defined vertical sensitivity is
identified in
format.
REMARKS
Rounding the value in
Rounding range
Set vertical
sensitivity
v_gain <= 2E-3
2mV/div
2E-3 < v_gain <= 5E-3 5mV/div
: :
5 < v_gain
10V/div
When the probe ratio is 10:1, 20:1, 100:1, 200:1, 1000:1 or 2000:1, the above
range is multiplied by 10, 100, 1000 or 2000.
• When the probe ratio is set to "AUTO", the above range is multiplied by 1,
10, or 100, depending on the detected probe ratio.
• Some value may be rounded to a permissible larger value.