Horizontal TRDL
Command/Query
DESCRIPTION The
TRDL command defines the horizontal position of the
trigger point in time. With the screen center defined
as zero, the time duration from this zero point to the
trigger point is set. The setting range is determined
by the timebase (TDIV). The TRDL? query returns the set
horizontal position of the trigger point.
COMMAND SYNTAX
TRDL
is also valid.
The range of
10ms/div)
10us/div)
500ps/div)
See Remarks for detail.
QUERY SYNTAX
TRDL?
RESPONSE FORMAT
used. The currently set trigger point is input here.
REMARKS
TRDL reports command-execution error (CME) when
For more detail information see below.
1. Setting Range
The setting range and rounding method for the TRDL command are defined as
follows, depending on the waveform acquisition and sweep time.
Trigger Mode
Sweep time
Setting range
50s/div to 10ms/div
-500s to
+5div
*
5ms/div to 10us/div
-5s to +5div
*
SINGLE, NORMAL or AUTO
5us/div to 500ps/div
-5ms to +5div
*
STOP
-
-500s to +750s
*
• The above division value in the setting range is calculated from the following
equation.
(Setting range upper limit) = +5div × (Sweep time)
• Any value exceeding the above range is rounded to the maximum or minimum
permissible value, whichever is nearest.
2. Set resolution
The resolution is changed according to the set sweep time. The resolution is
calculated from the following equation.