Wavejet 300a remote control manual – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual
Page 19
WaveJet 300A Remote Control Manual
917976 Rev A
19
is identified in
REMARKS WJ
324, WJ314, WJ322 and WJ312 set the bandwidth limiter
to 20 MHz when these scopes receive "BWL 100M".
Status *CLS
Command
DESCRIPTION
The *CLS command clears the entire status data
register.
COMMAND SYNTAX
*CLS
QUERY SYNTAX
RESPONSE FORMAT
Cursor CMSR?
Query
DESCRIPTION
The CMSR? query returns the currently displayed cursor
measurements.
COMMAND SYNTAX
QUERY SYNTAX
CMSR?
RESPONSE FORMAT
• Current cursor measurements are displayed in
• When the cursor mode is disabled or the data cannot
be measured, the query returns +9.91E+37.
• Seven types of cursor measurement results are output,
respectively separated by a comma, regardless of
whether cursor mode has been selected. For items not
covered by the measurement, the query returns
+9.91E+37.
• The table below shows the relationship between
cursor mode.
When MATH is FFT, M1:TRA is ON and CURM is VATH, CMSR?
returns frequency at the cursor of the FFT waveform for
REMARKS
When calculating the cursor measurement values of the
YT display Amplitude, Value at Cursor, and XY display
ch Amplitude in the model WJ352/332/322/312,