WaveJet 300A Remote Control Manual

917976 Rev A

47

Trigger TPTM

Command/Query

DESCRIPTION

The TPTM command defines the parameters for period
trigger (TTYP=PEIOD).
The TPTM? query returns the current parameters of
period trigger.

COMMAND SYNTAX

TPTM ,
:= {M_T, T_M}
M_T := Set period trigger interval time condition to
"m <= t".
T_M := Set period trigger interval time condition to
"t <= m".
:= Set "m" time value. Available range is 40 ns to
50.0 s. The format is
used, and the suffix is also valid.

QUERY SYNTAX

TPTM?

RESPONSE FORMAT

,
The currently set condition in . The format is
.
The currently set period time in . The format is NUMERIC RESPONSE DATA>.

REMARKS

TPTM command reports device-specific error (DDE) when
TTYP is not PERIOD.

Vertical TRA

Command/Query

DESCRIPTION

The TRA command enables or disables the trace
indication of the specified input channel. The TRA?
query returns the display status of the specified
trace.

COMMAND SYNTAX

:TRA
:={C1, C2, C3, C4, M1} (for WJ354/334/324/314)
:={C1, C2, M1} (for WJ352/332/322/312)
:={ON, OFF}

QUERY SYNTAX

:TRA?

RESPONSE FORMAT


The current trace display status is identified in
.

REMARKS

When XYDS is XY or XYTRG, C1 and C2 are not able to be
turned off, and C3 and C4 are not able to be turned on.