Wavejet 300a remote control manual – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual
Page 49
WaveJet 300A Remote Control Manual
917976 Rev A
49
(Delay value resolution) = (Sweep time) / 50
• When the specified resolution is not equal to the above value, it is rounded
to the permissible value of which the absolute value is smaller.
Acquisition *TRG
Command
DESCRIPTION
The *TRG command sets the trigger mode to "SINGLE" to
start single waveform acquisition.
COMMAND SYNTAX
*TRG
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
*TRG reports device-specific error (DDE) and is not
performed when XYDS is XY.
Acquisition TRMD
Command/Query
DESCRIPTION
The TRMD command selects the trigger mode. The TRMD?
query returns the currently selected trigger mode.
COMMAND SYNTAX
TRMD
AUTO := Set trigger mode to AUTO when scope is not in
roll mode operation. If roll mode is on and scope is
in roll mode operation range, AUTO initiates ENDLESS
roll acquisition.
NORMAL := Set trigger mode to NORMAL when scope is not
in roll mode operation. If roll mode is on and scope
is in roll mode operation range, NORMAL initiates
repeated TRIG'D roll acquisitions.
SINGLE := Set trigger mode to NORMAL when scope is not
in roll mode operation. If roll mode is on and scope
is in roll mode operation range, SINGLE initiates
single TRIG'D roll acquisition.
QUERY SYNTAX
TRMD?
RESPONSE FORMAT
Not in roll mode operation: AUTO, NORMAL, SINGLE or STOP
In roll mode operation: ENDLESS, TRIG'D, TRIG'D or STOP
REMARKS
TRMD SINGLE reports device-specific error (DDE) and is
not performed when XYDS is XY.