WaveJet 300A Series

46

917976 Rev A

Automatic Measurement TPRM

Command/Query

DESCRIPTION

The TPRM command sets the measurement conditions for
the automatic measurement item DELTAT set using the
MSEL command.
The TPRM? query returns the measurement conditions for
the automatic measurement item DELTAT set using the
MSEL command.

COMMAND SYNTAX

TPRM , , , , level>,
:={OFF, CH1, CH2, CH3, CH4, MATH}
(WJ354/334/324/314)
:={OFF, CH1, CH2, MATH}
(WJ352/332/322/312)
:= {RISE, FALL}
is set in
format. A voltage value in a range of –5 div to +5 div
from the center of the screen is set, assuming that the
ground level is 0 V.
:={CH1, CH2, CH3, CH4, MATH}
(WJ354/334/324/314)
:={CH1, CH2, MATH} (WJ352/332/322/312)
:= {RISE, FALL}
is set in
format. A voltage value in a range of –5 div to +5 div
from the center of the screen is set, assuming that the
ground level is 0 V.

QUERY SYNTAX

TPRM?

RESPONSE FORMAT

, , , , level>,
shows a voltage value in a range of –5 div
to +5 div set in format,
assuming that the ground level is 0 V.
shows a voltage value in a range of –5 div
to +5 div set in format,
assuming that the ground level is 0 V.

If is OFF, is returned as ch> is CH1.
If or is MATH and the MATH is FFT,
or becomes +9.91000E+37 that
shows it is invalid.

REMARKS

If or is set to CH3 or CH4 in models
WJ352/332/322/312, it is rounded to CH1.
TPRM command reports device-specific error (DDE) when
MSEL is not DELTAT.