WaveJet 300A Series
40
917976 Rev A
Save/Recall *SAV
Command
DESCRIPTION
The *SAV command saves the front panel setup data of
the oscilloscope to nonvolatile internal memory.
COMMAND SYNTAX
*SAV
0 := do nothing.
QUERY SYNTAX
RESPONSE FORMAT
Automatic Measurement SKLV
Command/Query
DESCRIPTION
The SKLV command defines the measurement conditions of
automatic measurement SKEW specified with the MSEL
command.
The SKLV? query returns the defined measurement
conditions.
COMMAND SYNTAX
SKLV
waveform is 100%. A value beyond the setting range is
rounded to a permissible value.
:={CH1, CH2} (WJ352/332/322/312)
waveform is 100%. A value beyond the setting range is
rounded to a permissible value.
QUERY SYNTAX
SKLV?
RESPONSE FORMAT
The parameter setting "Level1" of automatic
measurement SKEW is identified in
The parameter setting "Slope1" of automatic
measurement SKEW is identified in
The parameter setting "Source2" of automatic
measurement SKEW is identified in
The parameter setting "Level2" of automatic
measurement SKEW is identified in
The parameter setting "Slope2" of automatic
measurement SKEW is identified in
REMARKS
If
WJ352/332/322/312, it is rounded to CH1.
SKLV command reports device-specific error (DDE) when
MSEL is not SKEW.