Standard event status register, Standard event status enable register, Trigger event status register – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual
Page 14: Trigger event status enable register

WaveJet 300A Series
14
917976 Rev A
Standard Event Status Register
Table 2-10. Model of Standard Event Status Register
Bit Definition
Implemented?
7
Power On (PON)
Yes
6
User Request (URQ)
No (always zero)
5
Command ERROR (CME)
Yes
4
Execution ERROR (EXE)
Yes
3
Device-Specific Error (DDE)
Yes
2
Query ERROR (QYE)
Yes (only GPIB)
1
Request Control (RQC)
No (always zero)
0
Operation Complete (OPC)
Yes
Standard Event Status Enable Register
The Standard Event Status Enable Register allows one or more events in the Standard Event Status
Register to be reflected in the ESB summary-message bit.
Trigger Event Status Register
Table 2-11. Model of Trigger Event Status Register
Bit Definition
Implemented?
7
Always zero
No
6
Always zero
No
5
Always zero
No
4
Always zero
No
3
Always zero
No
2
Always zero
No
1
Always zero
No
0
Single Wave Enable (SWE)
Yes
Trigger Event Status Enable Register
The Trigger Event Status Enable Register allows one or more events in the Standard Event Status Register
to be reflected in the TESB summary-message bit.
2.22 For each command, a statement describing whether it is overlapped or sequential
Overlapped commands are not provided.
2.23 For each command device, the documentation shall specify the functional criteria that are
met when an operation complete message is generated in response to that command.
• Overlapped commands are not provided.
• The operation is completed on receipt of the command.