Trigger TWTM

Command/Query

DESCRIPTION The

TWTM command defines the parameters for pulse width

trigger (TTYP=PULSE).
The TWTM? query returns the current parameters of pulse
width trigger.

COMMAND SYNTAX

TWTM ,,
:= {M_T, T_M, M_T_N, T_M_N_T}
M_T := Set pulse width trigger time condition to "m <=
t".
T_M := Set pulse width trigger time condition to "t <=
m".
M_T_N := Set pulse width trigger time condition to "m
<= t <= n".
T_M_N_T := Set pulse width trigger time condition to
"t <= m, n<=t".
:= Set "m" time value. Available range is 15ns to
50.0s. The format is
used, and the suffix is also valid.
:= Set "n" time value. Available range is 15ns to
50.0s. The format is
used, and the suffix is also valid.

QUERY SYNTAX

TWTM?

RESPONSE FORMAT

,,
The currently set condition in . The format is
.
The currently set pulse width lower time in . The
format is .
The currently set pulse width upper time in . The
format is .

REMARKS

TWTM command reports device-specific error (DDE) when
TTYP is not PULSE.
When condition M_T_N or T_M_N_T is used, time and
take following range.
(m) <= (n) <= (m * 500)
If specified time value (m or n, or both) is out of this
range, the value is rounded to fit the range.