WaveJet 300A Series
54
917976 Rev A
Trigger TWTM
Command/Query
DESCRIPTION The
TWTM command defines the parameters for pulse width
trigger (TTYP=PULSE).
The TWTM? query returns the current parameters of pulse
width trigger.
COMMAND SYNTAX
TWTM
M_T := Set pulse width trigger time condition to "m <=
t".
T_M := Set pulse width trigger time condition to "t <=
m".
M_T_N := Set pulse width trigger time condition to "m
<= t <= n".
T_M_N_T := Set pulse width trigger time condition to
"t <= m, n<=t".
50.0s. The
used, and the suffix is also valid.
50.0s. The
used, and the suffix is also valid.
QUERY SYNTAX
TWTM?
RESPONSE FORMAT
The currently set condition in
The currently set pulse width lower time in
format is
The currently set pulse width upper time in
format is
REMARKS
TWTM command reports device-specific error (DDE) when
TTYP is not PULSE.
When condition M_T_N or T_M_N_T is used, time
(m) <= (n) <= (m * 500)
If specified time value (m or n, or both) is out of this
range, the value is rounded to fit the range.