beautypg.com

Wavejet 300a remote control manual – Teledyne LeCroy WaveJet 300A Series Remote Control User Manual

Page 27

background image

WaveJet 300A Remote Control Manual

917976 Rev A

27

Data transfer

DTSTUP

Command/Query

DESCRIPTION

The DTSTUP command writes setup data for the product.
The DTSTUP? query reads the setup data (the setup data
found when the query is issued) from the product.

COMMAND SYNTAX

DTSTUP#8
DTSTUP command, unlike other commands, requires two
steps.
Step 1: Send DTSTUP command without parameter. Then the
scope becomes ready to receive setup data block.
Step 2: Send setup data block in
#8 format. The contents of
consists of an 8-digit ASCII string. This
string must represent the number of bytes to be entered
into . The byte length should always be
a sign-free, 8-digit integer, and zero must not be
deleted.
The data transferred by the DTSTUP? query should be
found in .

QUERY SYNTAX

DTSTUP?

RESPONSE FORMAT

#8
・#8 shows that the response message is described in
the format of Data> and the content of the following
is a sign-free, 8-digit integer.
・The contents of represents the number
of bytes to be entered into . The byte
length should always be a sign-free, 8-digit integer
including zero.
・The contents of represent the
product's setup data in binary code.

REMARKS

This command does not support the multi-command.

Data transfer

DTWAVE?

Query

DESCRIPTION

The DTWAVE? query reads the waveform data.

COMMAND SYNTAX

QUERY SYNTAX

DTWAVE?

RESPONSE FORMAT

REMARKS

This command does not support the multi-command.