WaveJet 300A Remote Control Manual

917976 Rev A

53

Trigger TVSET

Command/Query

DESCRIPTION The

TVSET command sets field sequence, field number and

line number for TV trigger.
The TVSET query returns the currently set field
sequence, field number, and line number.

COMMAND SYNTAX

TVSET ,,
:= {1, 2, 4, 8} This specifies the field sequence
to 1, 2, 4 or 8.
:= 0, 1 to 8 (Specify 0 to selects "Any Fields")
:= 1 to 263 (When TVSTD is NTSC and is
odd number)
:= 1 to 262 (When TVSTD is NTSC and
is even number)
:= 1 to 313 (When TVSTD is PAL and is
odd number)
:= 314 to 625 (When TVSTD is PAL and
is even number)
:= 0 (Specify 0 to selects "Any Lines")

QUERY SYNTAX

TVSET?

RESPONSE FORMAT

,,

REMARKS

TVSET command reports device-specific error (DDE) when
TTYP is not TV.
Line number selection is automatically set as follows.
NTSC:263/262

PAL:625

Trigger TVSTD

Command/Query

DESCRIPTION The

TVSTD command selects the TV standard for TV trigger

(TTYP=TV).
The TVSTD? query returns the currently selected TV
standard.

COMMAND SYNTAX

TVSTD
:= {NTSC, PAL, CUSTOM}

QUERY SYNTAX

TVSTD?

RESPONSE FORMAT

REMARKS

TVSTD command reports device-specific error (DDE) when
TTYP is not TV.