Displays containing masks, Dual display grid selection – Teledyne LeCroy WavePro 7 Zi_Zi-A User Manual
Page 86
![background image](/manuals/353448/86/background.png)
WavePro 7 Zi/Zi-A Oscilloscopes
78
922137-00 Rev A
Note: If you have more than one waveform displayed on only one
grid, a second grid automatically opens when you select Next Grid.
Displays Containing Masks
Since masks are pixel-based display renderings, you can think of them as
part of the Display feature set on your instrument. Masks and mask testing
are applied to waveforms as part of Pass/Fail functionality accessed from
Analysis → Pass/Fail from the menu bar.
Note: For more detailed information about Masks and other Pass/Fail
functionality, such as removing a mask from your display, refer to Mask
Testing (on page 114).
Dual Display Grid Selection
Make a dual grid display selection for showing on an external display (once
properly connected to the I/O panel) by selecting Display → Display
Setup.... Numbered labels on this image correspond with the following
steps.
Note: First-time setup may show a different Display tab. Fix this by
touching File → Exit on the menu bar and re-launching the oscilloscope
software using the Start DSO shortcut on the desktop.